Abstract

Objectives: The aim of this study was to design an enzyme immunoassay based on a modified ELISA with high sensitivity for detecting digoxin.
Methods: The first step in development involved the conjugation of digoxin to the HRP (Horse Radish Peroxidase) enzyme via sodium metaperiodate oxidation. Surface modification, and thus assay modification, was achieved by the covalent immobilization of an anti-digoxin monoclonal antibody on a functional plate using 3-ATPES (3-aminopropyltriethoxysilane).
Results: The developed ELISA demonstrated superior sensitivity (0.026μg/ml) and lower variability in measurements repeated throughout a day, compared to a conventional ELISA (0.051μg/ml). This assay detected the exact amount of digoxin. The sensitivity and specificity of the modified ELISA surpassed other methods, and measurements were performed within a few hours. 

Conclusion: An efficient ELISA kit was produced, characterized by its affordability, ease of learning, and absence of a hand-washing step.

 

1. Cheng, J.W. and I. Rybak, Use of digoxin for heart failure and atrial fibrillation in elderly patients. The American journal of geriatric pharmacotherapy, 2010. 8(5): p. 419-427.
2. Castiglione, V., A. Aimo, G. Vergaro, L. Saccaro, C. Passino, and M. Emdin, Biomarkers for the diagnosis and management of heart failure. Heart failure reviews, 2022: p. 1-19.
3. Eichhorn, E.J. and M. Gheorghiade, Digoxin—new perspective on an old drug. New England Journal of Medicine, 2002. 347(18): p. 1394-1395.
4. Rich, M.W., F. McSherry, W.O. Williford, S. Yusuf, and D.I. Group, Effect of age on mortality, hospitalizations and response to digoxin in patients with heart failure: the DIG study. Journal of the American College of Cardiology, 2001. 38(3): p. 806-813.
5. Blažek, D. and V. Celer, The production and application of single-chain antibody fragments. Folia microbiologica, 2003. 48(5): p. 687-698.
6. Steinhauer, C., C. Wingren, A.-C.M. Hager, and C.A. Borrebaeck, Single framework recombinant antibody fragments designed for protein chip applications. Biotechniques, 2002. 33(S6): p. S38-S45.
7. Corso, C.D., A. Dickherber, and W.D. Hunt, An investigation of antibody immobilization methods employing organosilanes on planar ZnO surfaces for biosensor applications. Biosensors and Bioelectronics, 2008. 24(4): p. 805-811.
8. Das, R.D., S. Maji, S. Das, and C. RoyChaudhuri, Optimization of covalent antibody immobilization on macroporous silicon solid supports. Applied Surface Science, 2010. 256(20): p. 5867-5875.
9. Darain, F., K.L. Gan, and S.C. Tjin, Antibody immobilization on to polystyrene substrate—on-chip immunoassay for horse IgG based on fluorescence. Biomedical microdevices, 2009. 11: p. 653-661.
10. Dixit, C.K., S.K. Vashist, F.T. O’Neill, B. O’Reilly, B.D. MacCraith, and R. O’Kennedy, Development of a high sensitivity rapid sandwich ELISA procedure and its comparison with the conventional approach. Analytical chemistry, 2010. 82(16): p. 7049-7052.
11. Malakaneh, M., M. Rasaee, F. Rahbarizadeh, R. Madani, M. Forozandeh, K. Khabiri, et al., Characterization of a monoclonal antibody against neopterin using an enzyme-linked immunosorbent assay with penicillinase as label. Hybridoma, 2001. 20(2): p. 117-121.
12. Di Lorenzo, D., G. Roggeri, C. Iacobello, S. Belloli, M. Gion, D. Za Va, et al., Evaluation of a radioreceptor assay to assess exogenous estrogen activity in serum of patients with breast cancer. The International journal of biological markers, 1991. 6(3): p. 151-158.
13. Camara, P.D., K. Velletri, M. Krupski, M. Rosner, and W.C. Griffiths, Evaluation of the Boehringer Mannheim ES 300 immunoassay analyzer and comparison with enzyme immunoassay, fluorescence polarization immunoassay, and radioimmunoassay methods. Clinical biochemistry, 1992. 25(4): p. 251-254.
Files
IssueVol 1 No 3 (2023) QRcode
SectionOriginal Articles
DOI https://doi.org/10.18502/abi.v1i3.14550
Keywords
ELISA Surface modification Covalent bindin

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Enayati H, Mohammadnejad J, Nikfarjam A. Development of enzyme-linked immunosorbent assay (ELISA) based on covalent immobilization of antibody on plate for measurement of digoxin. ABI. 2023;1(3):139-144.